Coupling paste on Leeb hardness testers: when and how

A light part moves with the impact and the hardness looks lower. When to couple it to a base, how to do it properly and when the answer is something else.

Short answer

  • Coupling paste is used when the part is too light to measure by Leeb as it is: it is bonded to a solid base so it doesn't vibrate with the impact.
  • With the D device, below 2 kg on a rigid support (5 kg unsupported) you need to couple; once coupled, a 0.1 kg part 3 mm thick is enough.
  • If the part is even smaller, thinner or has no flat face to couple, the method is UCI.

The problem: the part moves with the impact

The symptom: on small parts, Leeb readings come out lower than expected and widely scattered.

Why it happens: Leeb measures how much the impact body rebounds. If the part is light, it moves or vibrates with the impact and takes some of the energy: the rebound is lower and the hardness looks lower than it is.

How it is solved: by making the part behave as if it were solid. Coupling paste bonds it to a heavy base, with no air in between, and the assembly absorbs the impact as a single piece.

Light part measured by Leeb: without paste it moves and vibrates with the impact and the rebound is lower; coupled with paste to a solid base, the assembly responds as a single piece. impactreboundimpactreboundpasteno pastewith pastethe part vibratesa single block
Without paste, the light part moves with the impact and takes some of the energy: the rebound drops and the hardness looks lower. With a thin layer of paste and no air, part and base absorb the impact together.

When it is needed: the minimum mass

The manufacturer sets the minimum part mass by impact device and how the part is supported:

Minimum part mass by Leeb device
DeviceNo supportRigid supportCoupledMin. thickness coupled
D, DC, DL, E5 kg2 kg0.1 kg3 mm
C1.5 kg0.5 kg0.02 kg1 mm
G15 kg5 kg0.5 kg10 mm

If the part meets the "no support" column, it needs nothing. If it only meets "on a rigid support", rest it on a solid, stable surface. Below that, couple it with paste.

If the part has a hardened layer (case-hardened, nitrided, induction-hardened), that layer must also have a minimum thickness: 0.8 mm with D, DC, DL and E devices, and 0.2 mm with C.

Which parts usually need it

Those that weigh little but have a flat face to couple:

  • Test pieces and samples cut for the lab or for incoming inspection.
  • Flat bars, plates and offcuts of a few kilos.
  • Small forged or machined parts: small flanges, tooling, die parts, large fasteners with a flat face.
  • Loose components checked before being assembled into a heavier unit.

Large, heavy parts don't need it, nor do parts already mounted and firmly held in a rigid structure, which acts as the support. And it doesn't work for parts without a flat face: there, the method is UCI (see below).

Coupling paste for hardness testers is supplied in a 50 ml tub, as an accessory for QH hardness testers.

How to couple, step by step

  1. Find a solid base: heavy, flat and clean, such as a thick steel block or plate.
  2. Check the part has a flat face to rest on; both contact surfaces must be flat and clean.
  3. Apply a thin layer of paste to the part's contact face.
  4. Press the part onto the base with a slight twisting motion, to spread the paste and push out the air. There must be no play or rocking.
  5. Measure at the centre of the part, with the device perpendicular, and take at least three readings.
dmq coupling paste for hardness testers
Coupling paste to bond light parts to a solid base.

When paste isn't the answer

  • Part below the minimum coupled mass or thickness (0.1 kg and 3 mm with the D device).
  • No flat face that can be coupled: curved, irregular or assembled parts.
  • An installed part that can't be moved onto a base.

In those cases, the method is UCI: the probe applies a load instead of an impact, needs no paste and only requires around 0.3 kg and 5 mm of thickness. See the guide Leeb or UCI.

Source: DEMEQ, QH5 and QH5 U hardness tester user manuals (minimum mass and thickness by device and probe).

Hardness testers for large and small parts

Leeb with coupling paste for light parts, or straight to UCI.

Other guides

If your question is a different one, it may already be answered in one of these. View all guides

Measuring hardness on site

The five problems that most distort a portable hardness reading and how to solve them.

Read the guide

Leeb or UCI

Which hardness method to choose depending on the size, mass and surface of the part.

Read the guide

Hardness conversion chart

HRC, HV and HB for steel, and which scales the Leeb value converts to by material.

Read the guide